National Research Council of Italy, Institute of Sciences of Food Productions (CNR-ISPA), Bari (Italy)
Elisa Santovito, Ali Abdallah, Valeria Ferraro

 

Background

Real-time monitoring of microbial oxygen consumption provides critical insights into metabolic activity, antibiotic susceptibility, and industrial bioprocess performance. The FirePlate-O2 system from PyroScience integrates optical dissolved‐oxygen sensors into standard 96-well microplates (OXMWP-96R), enabling continuous phase-angle (dphi), dissolved-oxygen (% air saturation), and signal-intensity (mV) readouts. However, the choice of well cover—whether a mineral oil overlay, standard plastic lid, or optical sealing film—can significantly influence oxygen diffusion and measurement accuracy. We compared three common sealing strategies to determine their impact on oxygen‐consumption kinetics during Staphylococcus aureus growth at three different cell densities. By identifying the optimal covering method, we aim to improve reproducibility and reliability in high-throughput respiratory assays.

 

Methods/Set-Up/Sample

The basic idea of this assay is to correlate the time, which is needed to reach 50% oxygen consumption in each well with the initial concentration of a microbial contamination of a sample. The 96-well oxygen sensor microplate with round bottom OXMWP-96R, each containing an embedded oxygen sensor at the bottom of every well, were inoculated with S. aureus suspensions at three initial concentrations: 10⁸, 10⁵, and 10³ CFU/mL. For each inoculum level, triplicate wells were assigned to one of three sealing approaches: (1) a 50 µL overlay of sterile mineral oil (MO) to minimize evaporation and gas exchange; (2) the manufacturer-supplied standard plastic lid (L), rested loosely on the plate to allow minimal gas exchange; or (3) an adhesive optical polyester acrylate sealing film  (OF). Negative-control wells (medium only) received identical covering. Plates were covered with the black lid and incubated at 35 °C in the FirePlate-O2 reader, which recorded dphi, dissolved O₂ (% air sat.), and signal (mV) every 10 minutes over a 12-hour period. Calibration was performed against air-saturated water (100 % O₂) and deoxygenated water (0 % O₂, using OXCAL Calibration Capsules) to ensure accurate sensor response.

FIG 1. FirePlate-O2 System Setup and Well‐Cover Comparisons. (A) FirePlate-O2 reader inside a temperature‐controlled incubation chamber, illustrating the overall instrument environment. (B) Close‐up of the FirePlate’s sensor block seated in the reader. (C) Top‐down view of a 96‐well Oxygen Sensor Microplate with integrated oxygen‐sensing spots (green), ready for measurement. (D) Comparison of three well‐covering strategies: (left) wells overlaid with sterile mineral oil, (center) wells covered by the standard plastic lid, and (right) wells sealed with adhesive optical film. Each yellow‐outlined panel highlights the uniformity of covers across a full row of 12 wells.

Data Aquisition and Observations

Figure 2 illustrates how the three well‐covering methods—mineral oil (MO, red), standard lid (L, green), and optical film (OF, blue)—affect oxygen‐consumption kinetics and reproducibility at high (10⁸ CFU/mL), intermediate (10⁵ CFU/mL), and low (10³ CFU/mL) S. aureus densities. It is clearly evident in Figure 2 that read-out of phase angle dphi or oxygen level are the superior method compared to read-out of signal intensity due to the enhanced reproducibility.

At 10⁸ CFU/mL and 10⁵ CFU/mL, all covering methods drive a rapid increase in dphi (indicating fast O₂ depletion), reaching their plateaus around 1 hour and ~4–5 hours respectively. At 10³ CFU/mL, MO‐covered wells and lid show virtually no dphi increase until after 5 hours, whereas film begin rising around 4.5 hours. Dissolved O₂ percentages mirror these patterns: OF-covered wells drop below 5 % O₂ earliest, whereas MO and L follow ~30–45 minutes later (most pronounced at low density). For each inoculum, OF wells achieve maximum intensity (indicative of near‐zero O₂) first, L next, and MO last.

Throughout the rapid‐depletion phases, OF yield the smallest SDs—tightest, most reproducible readings. L covered wells shows moderate SD peaks (roughly halfway between OF and MO), reflecting slightly more well‐to‐well variability.

MO wells exhibit the largest SD peaks across all inoculums, especially at low density, indicating uneven oil layering or inconsistent diffusion barriers.

In sum, standard OF allows the fastest, most consistent detection of oxygen depletion; L introduces a moderate delay and intermediate variability; MO creates the greatest lag in O₂ consumption and highest variability, particularly when cell densities are low.

FIG 2. Impact of Well Covers on Fireplate Oxygen Readouts at Three S. aureus Inocula.

Top row shows mean phase‐angle (dphi, °), dissolved O₂ (% air saturation), and signal‐intensity (mV) versus time for wells sealed with mineral oil (MO, red), standard lid (L, green), or optical film (OF, blue) at 10⁸, 10⁵, and 10³ CFU/mL. Higher inocula (10⁸, leftmost inflections) deplete O₂ fastest, followed by intermediate (10⁵) and low (10³) densities; MO consistently delays O₂ depletion relative to L and OF. Bottom row shows corresponding standard deviations: L (green) has the lowest variability, OF (blue) is intermediate, and MO (red) exhibits the highest well‐to‐well variation, especially during rapid oxygen‐consumption phases.

Conclusion from Data and Application

Optical films consistently produce the most reproducible oxygen-consumption profiles across all S. aureus inoculum levels, with minimal variability and rapid oxygen depletion. Mineral oil overlays lead to the slowest depletion and highest inter-well variability, rendering them suboptimal when precise early-phase respiratory kinetics are essential. Accordingly, we recommend optical film for routine FirePlate-O2 assays requiring high fidelity in dissolved O₂ measurements. Lid only can be employed when a controlled delay in O₂ depletion is beneficial, and mineral oil should be reserved for experiments prioritizing evaporation control over real-time oxygen accuracy.